Operation
Agilent 55B AA Spectrometer User’s Guide 89
3 Press the Results key.
4 Ensure the diluent tubing is connected to the Mariott vessel.
5 Place the pump tubing in the appropriate sample and press the Read key.
Results
The solution results are displayed on the Results page - see Page 64.
Print the results
You can print the solution results to an external printer, using the RS-232 9-pin
cable to connect the instrument to your PC or printer.
Figure 40. RS-232 9-pin socket located on the rear of the instrument
If the printer is not a serial printer, use a serial to parallel converter.
To print the results:
1 Go to Options page > Printer > Text .
2 Set the required ‘Baud rate’ and ‘Lines per page’.
3 Ensure the instrument is connected to the printer.
4 Press the Read key to measure the current solution.
As each solution is measured, the result will be sent to the printer.