Troubleshooting/Errors
Agilent 55B AA Spectrometer User’s Guide 137
Common problems
The following troubleshooting chart provides solutions to some common
problems you may encounter during routine AA analysis.
Please note that this list is not exhaustive.
Table 10. Common faults
a) Check gas supply pressure.
a) Check the burner type is correct for the selected
flame type and it is correctly installed.
a) Check liquid trap is full and float is free to move.
b) Check the operation of the liquid trap interlock in the
burner adjuster.
fault
a) Check pressure relief bung is correctly installed in the
spray chamber.
b) Check operation of pressure relief bung interlock.
gas flows
b) Check for blockages in gas supply lines.
a) Check O-rings in nebulizer block.
b) Check nebulizer is correctly assembled and adjusted.
c) Check for leaks.
no flame
a) Too much light in the sample compartment. Ensure
chimney and flame shields are fitted. Turn off nearby
lights.
b) Flame sensor faulty.
short
a) Low acetylene pressure, check supply.
b) Check that igniter solenoid operates correctly.
c) Check for blockage or carbon in igniter capillary.