Other Setup Tasks 5
To confirm the collision cell gas flow rate (6470A and 6495A/B only)
Agilent 6400 Series Triple Quad LC/MS System Installation Guide 99
9 For 6495A/B, get CCDAC_Setting_MRM_6495.m from the Agilent Support
Portal and load it.
As an alternative, you can set up source settings as shown in Figure 50.
Figure 50 Ion source settings for 6495A/B
10 For 6470, get CCDAC_Setting_MRM_6470.m from the Agilent Support
Portal and load it.
As an alternative, you can set up source settings as shown in Figure 50.
Tabl e 1 0 Minimum recommended tune abundances at gain = 10
6
in NEGATIVE Mode
(HED = 18 kV)
6495A/B 6470A
m/z MS1 MS2 MS1 MS2
113 6.E+05 6.E+05 7.E+04 4.E+04
302 2.E+06 2.E+06 2.E+05 2.E+05
602 2.E+06 1.E+06 1.E+05 1.E+05
1034 4.E+05 4.E+05 3.E+05 2.E+05
1634 9.E+05 5.E+05 5.E+05 4.E+05
2234 3.E+05 2.E+05 2.E+05 6.E+04