General Maintenance 4
Troubleshooting and Maintenance Manual 87
6 Place a column nut with a blank ferrule on the end of the interface. This
will help keep contamination out of the MS.
7 Disconnect the GC/MS interface cable. Disconnecting the cable with the
GC on can cause a fault condition.
8 Continue to move the MS until you have access to the part requiring
maintenance.
Reconnection
1 Position the MS so the end of the GC/MS interface is near the GC
(Figure 6).
2 Reconnect the GC/MS interface cable.
Figure 6 Separating/connecting the MS and GC
GC/MS interface
Interface cable