Setting Up the 708-DS / 709-DS 3
708-DS / 709-DS Operator’s Manual 27
Figure 6 Raising the Stability Feet
3 Using the 90º Allen key in the top of the front level-adjustment
screws, adjust the screws to achieve left-to-right level within
tolerance. It may be necessary to loosen the nut(s) beneath the base
plate to allow for adjustment.
4 Using the 90º Allen key in the top of the rear level-adjustment screw,
adjust the screw to achieve front-to-back level within tolerance. It
may be necessary to lift the heater / circulator to gain access to this
screw.
5 Verify the level of the dissolution apparatus with a bubble or digital
level.
6 Once the unit is leveled, use the open-end wrench to tighten the
nut(s) below the base plate.
7 Lower the side stability feet located toward the back on either side of
the base plate until they touch the laboratory bench. Do not
over-tighten to ensure level is maintained.
8 Reinstall the black caps on all of the screws.