Operating the 708-DS / 709-DS 4
708-DS / 709-DS Operator’s Manual 87
Menu Screen - Calibration
Menu Screen - Calibration Screen 1
The Calibration screen allows you to fine-tune the height of the
manifold for Apparatus 1 (Baskets) and Apparatus 2 (Paddles). The
volumes shown change based on the installed vessel size set in
Instrument Setting Screen 2 (“Menu Screen - Instrument Settings
Screen 2” on page 84). The following vessel size ranges are
configurable:
Size Max Min
2L 2000 mL 1650 mL
1L 900 mL 500 mL
200 mL 200 mL 150 mL
100 mL 100 mL 80 mL