Operating the 708-DS / 709-DS 4
708-DS / 709-DS Operator’s Manual 49
 
Main Screen - Sample Temperatures
When you press Sample Temperature, the Sample Temperatures box 
displays. The current bath temperature is displayed in the Bath Probe 
region and the current vessel probe temperature is displayed in the 
Probe region. Initially, the temperature at all vessel locations is 
displayed as "----".
If the apparatus is in idle mode or running a manual method, press the 
area over the displayed RPM to quick-set the spindle speed. Upon 
pressing this area, the numeric data entry screen displays. Enter a valid 
number to change the RPM and press Ok to return to the Main screen. 
The spindle speed changes only if the spindle was previously running. If 
the spindle was idle, it starts.
Similarly, if you press the area over the displayed temperature, you can 
quick-set the bath temperature. Upon pressing this area, the numeric 
data entry screen displays. Enter a valid number to set the bath 
temperature setpoint. Press Ok to return to the Main screen.
Sample Temperatures is available only if Vessel Probe is selected in the 
manufacturing configuration.