104 708-DS / 709-DS Operator’s Manual
4 Operating the 708-DS / 709-DS
 
Warning Screens (Yellow)
Table 1 6 Warning Screens (Yellow)
Instrument Calibration Due
This Service Due warning displays when the instrument is due for 
calibration (“Menu Screen - Calibration Screen 1” on page 87). Press 
Ok to clear the warning.
Instrument Preventative Maintenance Due
This Service Due warning displays when the instrument is due for 
preventative maintenance (“Preventive Maintenance” on page 116). 
Press Ok to clear the warning.
Temperature Probe Dwell Time Warning
This Method Warning screen displays when an Auto Sampling Manifold 
is installed and enabled (“Method Properties Screen 3” on page 72). If 
the sum of the Manifold Down Duration and Manifold Lead Time 
(“Method Properties Screen 3” on page 72) are less than 30 seconds, this 
screen displays. Press Ok to clear the warning.
Invalid Initial Temperature Warning
This Method Warning screen displays only when Apparatus 1 (basket) 
or Apparatus 6 (rotating cylinder) are installed and a Bath Temperature 
Start is selected from the Start Options screen (“Method - Start 
Options” on page 52). Press Ok to clear the warning and start the 
method without the initial temperature being taken. Press Cancel to 
abort the run and return to the Main screen.