88 708-DS / 709-DS Operator’s Manual
4 Operating the 708-DS / 709-DS
 
Each of the gray boxes contains a numeric value. The value is the 
sampling location of the manifold as measured from its home position. 
These values range from 0 at home to 999 at the manifold's lowest 
position. If no values have ever been entered, the default values, 
protected in the firmware, are shown for each Apparatus. If previous 
values have been entered, these are displayed. Default values for a 1L 
system are shown in the screen shot.
To set a new position, press Manifold   or Manifold   to raise or 
lower the manifold. The manifold will move as long as the button is 
held. The current location of the manifold is displayed above 
Manifold  . Once the desired position is reached, press Save on the 
corresponding field. This updates the display for that field.
From the Paddles/Baskets Sampling Depth screen:
• Press Restore Defaults to replace all of the displayed values with 
the default values, which are protected in the firmware.
• Press Cancel to exit the routine without saving the values and 
return to the Menu screen.
• Press Finish to save all values to permanent memory and return to 
the Menu screen.
• Press Next to save all values to permanent memory and proceed 
to the next Calibration screen.