Tuna ble Laser S ou rces Error Messages
Agilent 8163A/B, 8164A/B & 8166A/B Mainframes, Sixth Edition 399
error message TCFS Amplifier Selftest Failed (error value)
description This test calibrates the TCFS A, B and M channel and verifies if all resulting
values are in limits
error value 25041 TCFS Hybrid amplifier offset could not be removed
25042 One of the channels A, B or M failed during calibration
action Please return the unit to your local service center for repair
error message TCFS Adj. Cicuit Selftest Failed (error value)
description Checks whether a signal between +15V and +21V can be selected at the 20V
reference with the DAC
error value 25050 TCFS 20V adjust is not possible
action Please return the unit to your local service center for repair
error message TCFS Shutter Selftest Failed (error value)
description This test verifies the shutter driver by applying a designated test pattern
error value 25060 TCFS shutter driver electronic fails
action Please return the unit to your local service center for repair