SCPI COMMAND SUMMARY
If this is
ON
, each time a frequency or p ower is changed,
CALibration:AM[:EXECute]
is attempted.
CALibration:AM[:EXECute]
When AM is on and the swept signal generator is in the CW or
manual mo de, the swept signal generator p erforms a mo dulator
calibration as long as p ower sweep is not active.
CALibration:PEAKing:AUTO ON|OFF|1|0
CALibration:PEAKing:AUTO?
Sets and queries the automatic peaking function. If
AUTO
is
ON
, then
a p eak is done at regular intervals automatically. After *RST, the
setting is
OFF
.
CALibration:PEAKing[:EXECute]
Peaks the SYTM.
CALibration:PMETer:DETector:INITiate? IDETector|DIODe
Initiates the sp ecied calibration. These calibrations require the use
of an external power measurement. Once initiated, the swept signal
generator sets up for the rst p oint to b e measured, and responds to
the query with the frequency at which the power is to be measured.
The parameters mean:
IDETector
Initiates a calibration of the internal detector logger
breakpoints and osets.
DIODe
Initiates a calibration of an external detector's logger
breakpoints and osets.
CALibration:PMETer:DETector:NEXT? <num>[lvl suffix]
The parameter is the measured p ower that is currently pro duced by
the swept signal generator. You must supply this parameter after
measuring the p ower using an external p ower meter. The query
response is issued after the swept signal generator pro cesses the
supplied parameter and settles on the next p oint to b e measured.
The query resp onse is:
>
0 The frequency [in Hz] that is currently pro duced.
0 The calibration is complete.
<
0 An error has o ccurred and the calibration is ab orted.
CALibration:PMETer:FLATness:INITiate?
USER|DIODe|PMETer|MMHead
Initiates the sp ecied calibration. These calibrations require the use
of an external power measurement. Once initiated, the swept signal
generator sets up for the rst p oint to b e measured, and responds to
the query with the frequency at which the power is to be measured.
The parameters mean :
USER
Initiates a calibration at all of the user atness p oin
ts.
S-28 Operating and Programming Reference

Table of Contents

Other manuals for Agilent Technologies 8360 B Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8360 B Series and is the answer not in the manual?

Agilent Technologies 8360 B Series Specifications

General IconGeneral
BrandAgilent Technologies
Model8360 B Series
CategoryPortable Generator
LanguageEnglish

Related product manuals