SCPI COMMAND SUMMARY
FM:COUPling AC|DC
FM:COUPling?
Sets and queries the FM input coupling mo de.
The *RST value is AC.
FM[:DEViation] <num>[freq suffix>]|MAXimum|MINimum
FM[:DEViation]? [MAXimum|MINimum]
Sets and queries the peak FM deviation (in Hz) when the internal
FM generator is used. After *RST, the value is 1 MHz.
FM:FILTer:HPASs <num>[freq suffix'|MAXimum|MINimum
FM:FILTer:HPASs? [MAXimum|MINimum]
Sets and queries the FM AC bandwidth. There are only two
positions to the bandwidth,
<
20 Hz and
>
100 kHz, but any
numeric is accepted. The value is compared to 1 kHz and the correct
position is determined (
>
1 kHz sets the p osition to 100 kHz and
1 kHz sets the p osition to 20 Hz). After *RST, the value is
100 kHz.
FM:INTernal:FREQuency <num>[freq suffix]|MAXimum|MINimum
FM:INTernal:FREQuency? [MAXimum|MINimum]
Sets and queries the frequency (in Hz) of the in
ternal FM source.
After *RST, the value is 1 MHz.
FM:INTernal:FUNCtion SINusoid|SQUare|TRIangle|RAMP|NOISe
FM:INTernal:FUNCtion?
Sets and queries the waveform of the internal FM source.
After *RST, the value is SINusoid.
FM:SOURce INTernal|EXTernal
FM:SOURce?
Sets and queries the source of the FM mo dulating signal.
After *RST, the value is EXTernal.
FM:SENSitivity <num>[freq suffix/V]|MAXimum|MINimum
FM:SENSitivity? [MAXimum|MINimum]
Sets and queries the FM Input sensitivity.
The *RST value is MAX (10 MHz/V) .
FM:STATe ON|OFF|1|0
FM:STATe?
Sets and queries the FM mo dulation state.
After *RST, the value is OFF.