SCPI COMMAND SUMMARY
Sets and queries the type of preset to execute when the
SYSTem:PRESet[:EXECute]
command is given. Factory preset
defaults all values to factory sp ecied values. User dened preset
defaults all values to a sp ecied state of the swept signal generator
that you havesaved with
SYSTem:PRESet:SAVE
.
SYSTem:SECurity:COUNt <num>|MINimum|MAXimum
SYSTem:SECurity:COUNt? [MINimum|MAXimum]
Sets and queries the number of times to clear memory when the value
of SYSTem:SECurity[:STATe] changes from ON to OFF.
*RST value is 1.
SYSTem:SECurity[:STATe] ON|OFF|1|0
SYSTem:SECurity[:STATe]?
Controls the security mo de of the swept signal generator. When
in secure mo de, any display annunciators that have b een disabled
cannot b e re-enabled without destroying certain stored information.
This value is not aected by *RST. When you change the value
from ON to OFF, everything except calibration data is initialized or
destroyed. In particular, data in instrument state and all save/recall
registers are destroyed.
SYSTem:VERSion?
This query returns a formatted numeric value corresp onding to the
SCPI version number to which the swept signal generator complies.
The response has the form YYYY.V where the Ys represen
t the year
version (i.e. 1990) and the V represents an approved revision number
for that year. This is a query only and therefore do es not ha
vean
asso ciated *RST state.
TRIGger[:IMMediate]
Causes the trigger event to o ccur regardless of other settings in the
subsystem. This event do es not aect any other settings in this
subsystem.
This command has no eect unless the swept signal generator is in
the
wait for trig
state. If the swept signal generator is in the
wait for
trig
state, it p erforms its trigger action. This is an ev
ent and has no
*RST condition.
TRIGger:ODELay <num>[time suffix]|MAXimum|MINimum
TRIGger:ODELay? [MAXimum|MINimum]
Sets and queries the trigger output delay, the time between when
the source is settled, (when Bit 1 of the Standard Op eration Status
Register makes a negative transition), and the trigger out signal is
sent.
S-54 Operating and Programming Reference

Table of Contents

Other manuals for Agilent Technologies 8360 B Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8360 B Series and is the answer not in the manual?

Agilent Technologies 8360 B Series Specifications

General IconGeneral
BrandAgilent Technologies
Model8360 B Series
CategoryPortable Generator
LanguageEnglish

Related product manuals