Figure 1-20.
Creating Arbitrarily Spaced Frequency-Correction Pairs
inaSwept mm-wave Environment
For this example, refer to menu map 5, POWER.
1. The equipment setup shown in Figure 1-20 assumes that
you have followed the steps necessary to correctly level the
conguration. If you have questions ab out external leveling, refer
to \Externally Leveling the Swept Signal Generator".
Setup Power Meter
2. Zero and calibrate the p ower meter/sensor.
3. Connect the power sensor to test p ort.
4. Enter and store in the p ower meter, the p ower sensor's cal factors
for correction frequencies to be used.
1-40 Getting Started Advanced