If a GPIB error message is displayed verify that the interface
connections are correct. Check the GPIB address of the p ower
meter and ensure that it is the same address the swept signal
generator is using (address 13 is assumed). Refer to the menu
map 8, System, for the key sequence necessary to reach softkey
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Meter Adrs
.
9. When the op eration is complete, (a message is displayed)
disconnect the p ower meter/sensor. The swept signal generator
has stored the comp ensation information in its memory and is
using it to calibrate the detector's output voltage relativeto
power.