Remote
Remote causes an instrumenttochange from lo cal control to
remote control. In remote control, the front panel keys are disabled
(except for the
4
LOCAL
5
key and the POWER switch), and the amber
REMOTE annunciator is lighted. The syntax is:
where the device selector is the address of the instrument app ended
to the GPIB p ort number. Typically, the GPIB p ort number is 7,
and the default address for the swept signal generator is 19, so the
device selector is 719. Some BASIC examples:
10 REMOTE 7
which prepares all GPIB instruments for remote op eration (although
nothing app ears to happen to the instrumen
ts until they are
addressed to talk), or
10 REMOTE 719
which aects the GPIB instrument lo cated at address 19, or
10 REMOTE 719, 721, 726, 715
which eects four instruments that have addresses 19, 21, 26, and 15.
Related statements used by some computers:
RESUME
Local Lockout
Local Lo ckout can b e used in conjunction with REMOTE to disable
the front panel
4
LOCAL
5
key. With the
4
LOCAL
5
key disabled, only the
controller (or a hard reset by the POWER switch) can restore lo cal
control. The syntax is:
A BASIC example:
10 REMOTE 719
20 LOCAL LOCKOUT 7
1-58 Getting Started Programming

Table of Contents

Other manuals for Agilent Technologies 8360 B Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8360 B Series and is the answer not in the manual?

Agilent Technologies 8360 B Series Specifications

General IconGeneral
BrandAgilent Technologies
Model8360 B Series
CategoryPortable Generator
LanguageEnglish

Related product manuals