Programmer’s Guide 13-9
SCPI Error Messages
Execution Errors
Table 13-2 SCPI Execution Errors
Error
Number
Error Description
200 Execution error — This is the generic syntax error for devices that cannot
detect more specific errors. This code indicates only that an Execution Error
has occurred.
201 Invalid while in local — Indicates that a command is not executable while
the device is in local due to a hard local control; for example, a device with a
rotary switch receives a message which would change the switches state,
but the device is in local so the message can not be executed.
202 Settings lost due to rtl — Indicates that a setting associated with a hard
local control was lost when the device changed to LOCS from REMS or to
LWLS from RWLS.
203 Command protected — Indicates that a legal password-protected program
command or query could not be executed because the command was
disabled.
210 Trigger error
211 Trigger ignored — Indicates that a GET, *TRG, or triggering signal was
received and recognized by the device but was ignored because of device
timing considerations; for example, the device was not ready to respond.
1
212 Arm ignored — Indicates that an arming signal was received and
recognized by the device but was ignored.
213 Init ignored — Indicates that a request for a measurement initiation was
ignored as another measurement was already in progress.
214 Trigger deadlock — Indicates that the trigger source for the initiation of a
measurement is set to GET and subsequent measurement query is received.
The measurement cannot be started until a GET is received, but the GET
would cause an INTERRUPTED error.
215s Arm deadlock — Indicates that the arm source for the initiation of a
measurement is set to GET and subsequent measurement query is received.
The measurement cannot be started until a GET is received, but the GET
would cause an INTERRUPTED error.

Table of Contents

Other manuals for Agilent Technologies 8712ET

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8712ET and is the answer not in the manual?

Agilent Technologies 8712ET Specifications

General IconGeneral
BrandAgilent Technologies
Model8712ET
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals