Programmer’s Guide 5-25
Using Status Registers
The Analyzer's Status Register Sets
Settings for STATus:PRESet
Executing the STATus:PRESet command changes the settings in the
enable (ENAB), positive transition (PTR), and negative transition (NTR)
registers. The table below shows the settings after the command is
executed.
Table 5-1 Status Register States After PRESet Command
Register Set ENABle PTRansition NTRansition
STATus:DEVice all 0s all 1s all 0s
STATus:QUEStionable:LIMit all 1s all 1s all 0s
STATus:QUEStionable all 0s all 1s all 0s
STATus:OPERation:MEASuring all 1s all 0s all 1s
STATus:OPERation:AVERaging all 1s all 0s all 1s
STATus:OPERation all 0s all 1s all 0s