Programmer’s Guide 4-3
Data Types and Encoding
Data Types
Data Types
The analyzer uses a number of different data types during data
transfers. Data transfer occurs in response to a query. The data type
used is determined by the parameter being queried. Data types described
in this section are:
Numeric Data
Character Data
String Data
Expression Data
Block Data
Numeric Data
The analyzer returns three types of numeric data in response to queries:
NR1 data Integers (such as +1, 0, -1, 123, -12345). This is the
response type for boolean parameters as well as some
numeric parameters.
NR2 data Floating point numbers with an explicit decimal point
(such as 12.3, +1.234, -0.12345).
NR3 data Floating point numbers in scientific notation (such as
+1.23E+5, +123.4E-3, -456.789E+6).
Character Data
Character data consists of ASCII characters grouped together in
mnemonics that represent specific instrument settings (such as
MAXimum, MINimum or MLOGarithmic). The analyzer always returns the
short form of the mnemonic in upper-case alpha characters.

Table of Contents

Other manuals for Agilent Technologies 8712ET

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8712ET and is the answer not in the manual?

Agilent Technologies 8712ET Specifications

General IconGeneral
BrandAgilent Technologies
Model8712ET
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals