6-22 Programmer’s Guide
Trace Data Transfers
Internal Measurement Arrays
CH[1|2]BFWD and CH[1|2]RFWD. Raw data for AUX INPUT is not
available via GPIB. Use the corrected data array to access AUX INPUT
data.
Table 6-4 Raw Data Arrays
Ratio Calculations
These are performed if the selected measurement is a ratio (e.g. A/R or
B/R). This is simply a complex divide operation. If the selected
measurement is absolute (e.g. A or B), no operation is performed.
Error Correction
Error correction is performed next if correction is turned on. Error
correction removes repeatable systematic errors (stored in the error
coefficient arrays) from the raw arrays. The operations performed
depend on the selected measurement type.
Selected Measurement Raw Arrays
Transmission (B/R) B = CH[1|2]BFWD, R=CH[1|2]RFWD
Reflection (A/R) A = CH[1|2]AFWD, R= CH[1|2]RFWD
AA =CH[1|2]AFWD
BB =CH[1|2]BFWD
RR =CH[1|2]RFWD
Power (B*) B*= CH[1|2]BFWD
Conversion Loss (B*/R*) B*= CH[1|2]BFWD, R*= CH[1|2]RFWD
R* R*= CH[1|2]RFWD
AM Delay (Y/X) Y = CH[1|2]BFWD, X = CH[1|2]RFWD
XX= CH[1|2]RFWD
YY =CH[1|2]BFWD
Y/R* Y = CH[1|2]BFWD, R* = CH[1|2]RFWD
Y/X, X/Y Y = CH[1|2]BFWD, X = CH[1|2]RFWD

Table of Contents

Other manuals for Agilent Technologies 8712ET

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8712ET and is the answer not in the manual?

Agilent Technologies 8712ET Specifications

General IconGeneral
BrandAgilent Technologies
Model8712ET
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals