Programmer’s Guide 5-23
Using Status Registers
The Analyzer's Status Register Sets
Measuring Status Register Set
The Measuring Status register set monitors conditions in the analyzer's
measurement process.
Bits in the Measuring Status condition register are set to 1 under the
following conditions:
Channel 1
Measuring (bit 0) is set to 1 while the analyzer is collecting
measurement data on channel 1.
Channel 2
Measuring (bit 1) is set to 1 while the analyzer is collecting
measurement data on channel 2.
Averaging Status Register Set
The Averaging Status register set monitors conditions in the analyzer's
measurement process when the trace averaging function is in use.
Bits in the Averaging Status condition register are set to 1 under the
following conditions:
Measurement
Channel 1
Averaging (bit 0) is set to 1 while the analyzer is sweeping on
measurement channel 1 and the number of sweeps
completed (since “average restart”) is less than the
averaging factor.
Measurement
Channel 2
Averaging (bit 1) is set to 1 while the analyzer is sweeping on
measurement channel 2 and the number of sweeps
completed (since “average restart”) is less than the
averaging factor.

Table of Contents

Other manuals for Agilent Technologies 8712ET

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8712ET and is the answer not in the manual?

Agilent Technologies 8712ET Specifications

General IconGeneral
BrandAgilent Technologies
Model8712ET
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals