Programmer’s Guide 5-13
Using Status Registers
The Analyzer's Status Register Sets
Bits in the Status Byte register are set to 1 under the following
conditions:
Device Status Summary
(bit 2) is set to 1 when one or more enabled bits in the Device
Status event register are set to 1.
Questionable Status Summary
(bit 3) is set to 1 when one or more enabled bits in the
Questionable Status event register are set to 1.
Message Available
(bit 4) is set to 1 when the output queue contains a response
message.
Standard Event Status Summary
(bit 5) is set to 1 when one or more enabled bits in the Standard
Event Status event register are set to 1.
Master Summary Status
(bit 6, when read by *STB) is set to 1 when one or more enabled
bits in the Status Byte register are set to 1.
Request Service
(bit 6, when read by serial poll) is set to 1 by the service request
process (see “Generating a Service Request” on page 5-8).
Operational Status Summary
(bit 7) is set to 1 when one or more enabled bits in the
Operational Status event register are set to 1.

Table of Contents

Other manuals for Agilent Technologies 8712ET

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8712ET and is the answer not in the manual?

Agilent Technologies 8712ET Specifications

General IconGeneral
BrandAgilent Technologies
Model8712ET
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals