EasyManua.ls Logo

Agilent Technologies 8712ET

Agilent Technologies 8712ET
320 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Programmer’s Guide 5-13
Using Status Registers
The Analyzer's Status Register Sets
Bits in the Status Byte register are set to 1 under the following
conditions:
Device Status Summary
(bit 2) is set to 1 when one or more enabled bits in the Device
Status event register are set to 1.
Questionable Status Summary
(bit 3) is set to 1 when one or more enabled bits in the
Questionable Status event register are set to 1.
Message Available
(bit 4) is set to 1 when the output queue contains a response
message.
Standard Event Status Summary
(bit 5) is set to 1 when one or more enabled bits in the Standard
Event Status event register are set to 1.
Master Summary Status
(bit 6, when read by *STB) is set to 1 when one or more enabled
bits in the Status Byte register are set to 1.
Request Service
(bit 6, when read by serial poll) is set to 1 by the service request
process (see “Generating a Service Request” on page 5-8).
Operational Status Summary
(bit 7) is set to 1 when one or more enabled bits in the
Operational Status event register are set to 1.

Table of Contents

Other manuals for Agilent Technologies 8712ET

Related product manuals