Chapter 3 3-25
Adjustments and Correction Constants
Sampler Magnitude and Phase Correction Constants (Test 53)
A Channel Sampler
1. Access the first address by pressing
.
2. Enter the new value for the A sampler offset at the accessed address by pressing
[new value for A] .
3. Access the second address by pressing .
4. Enter the new value at the accessed address by pressing .
5. Press for the analyzer to use the new values.
6. Go to “Finish,” or to the next sampler to be adjusted.
B Channel Sampler
1. Access the first address by pressing
.
2. Enter the new value for the B sampler offset at the accessed address by pressing
[new value for B] .
3. Access the second address by pressing .
4. Enter the new value at the accessed address by pressing .
5. Press for the analyzer to use the new values.
6. Go to “Finish,” or to the next sampler to be adjusted.
R Channel Sampler
1. Access the first address by pressing
.
2. Enter the new value for the R sampler offset at the accessed address by pressing
[new value for R] .
3. Access the second address by pressing .
4. Enter the new value at the accessed address by pressing .
5. Press for the analyzer to use the new values.
6. Go to “Finish,” or to the next sampler to be adjusted.
Finish
1. Repeat the “Sampler Correction Constants Routine” on page 3-20.
2. If the analyzer continues to fail this adjustment routine, refer to Chapter 7 , “Source
Troubleshooting.”
System
1619001372 x1
x1
1619001373 x1
248 x1
Preset
System
x1
x1
1619001375 x1
248 x1
Preset
System
1619001376 x1
x1
1619001377 x1
128 x1
Preset