EasyManua.ls Logo

Agilent Technologies E4406A VSA Series - Page 22

Agilent Technologies E4406A VSA Series
124 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
22 Chapter2
Setting Up the EDGE (with GSM) Mode
Changing the Mode Setup
RF Sync Delay - In measurements that detect the GSM “T0”,
RF Sync Delay adjusts the “T0” point. This adjustment does not apply
if the
Burst Sync key is set to None,orifitissettoTraining Seq in the
Phase and Frequency Error measurement. The “T0” point is defined
as the time point of the transition from bit 13 to bit 14 of the
midamble training sequence for a given time slot.
Burst Search Threshold - Sets the threshold level used in the search
for EDGE (with GSM) bursts after data is acquired. This is a relative
level based on the peak “on” power.

Table of Contents

Other manuals for Agilent Technologies E4406A VSA Series

Related product manuals