Chapter 4 4-37
Troubleshooting
201 to 799: Device-Specific Errors
607 File Name Error
An invalid file name has been specified. Use filenames
with a maximum of 8 characters (letters and digits
only) and use a 3 character extension. Note that
lowercase and uppercase are perceived as the same.
This error will also occur if you attempt to delete a
nonexistent file.
612 File not found
The analyzer could not find the specified file.
613 Flash memory is full
The internal flash memory is full. Clear some space by
deleting unwanted files. You may also increase the flash
memory size by purchasing Option B72.
602 Floppy disk error
An unknown error has occurred while accessing the
floppy disk.
601 Floppy disk full
The floppy disk is full. Clear some space by deleting
unwanted files.
618 Illegal write access of Flash memory
Attempt to write to an unavailable area of internal
flash memory.
727 In <filename>: [DATA] header missing
This message indicates that the data section of a file
did not begin with the token [
DATA].
728 In <filename>, line <nnn>: separator missing
The [
HEADER] section of a file contains entries
requiring an equal (=) sign, such as <keyword> =
<value>. This message appears if the equal sign does
not appear on the line.

Table of Contents

Other manuals for Agilent Technologies E4408B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E4408B and is the answer not in the manual?

Agilent Technologies E4408B Specifications

General IconGeneral
BrandAgilent Technologies
ModelE4408B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals