Chapter 4 4-45
Troubleshooting
Greater than 1000: Personality Specific Error Messages
Greater than 1000:
Personality Specific Error Messages
An error detected with a number greater than 1000 indicates the
instrument has detected an error relating to an installed personality.
For more information on these errors, refer to the user’s guide for the
personality in use.
10219 Awaiting trigger
A trigger event from the selected trigger source has not
been detected.
10164 Band Measurement not defined for Out of Band.
User is attempting to monitor the band but has set the
frequency outside the band. Reset the band for the
particular standard for which you are testing or use the
channel setting which does not require a frequency to
be set. (
Meas Setup, Method (Channel)).
10228 Cannot correlate to input signal.
This error is normally generated because of one of the
following reasons: 1. There is no carrier signal.
2. Walsh channels other than the pilot are active.
3. There is some other modulation problem that will
prevent the measurement from being made.
This problem must be corrected before the
measurement can continue.
10163 Cannot find the Power vs Time Limits File.
The limit line definition file for the GSM standards has
been deleted. This message is displayed while the
Measure key is grayed out. Reinstall the GSM
measurement personality.
10166 Cannot update the list of cable types.
The cable file may have been moved or deleted
accidentally. Reinstall the GSM measurement
personality.

Table of Contents

Other manuals for Agilent Technologies E4408B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E4408B and is the answer not in the manual?

Agilent Technologies E4408B Specifications

General IconGeneral
BrandAgilent Technologies
ModelE4408B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals