EasyManuals Logo
Home>Agilent Technologies>Measuring Instruments>E4408B

Agilent Technologies E4408B User Manual

Agilent Technologies E4408B
321 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #69 background imageLoading...
Page #69 background image
Chapter 3 3-9
Functional Testing
Displayed Average Noise Level
Procedure (1.01 GHz to 1.5 GHz) E4401B and E4411B
1. Press the following keys on the analyzer:
Sweep, Sweep (Cont)
FREQUENCY, Start Freq, 1.0 GHz
Stop Freq, 1.5 GHz
BW/Avg, 1 MHz
Video BW Auto Man, 10 kHz
2. Press the following keys on the analyzer:
Single
BW/Avg, Average (On), 3, Enter
Single
Wait until AVG 3 is displayed to the left of the graticule (the
analyzer will take 3 sweeps, then stop).
3. Press Peak Search, and record the marker frequency in the Measured
Frequency column (c) in Table 3-1 for 1.01 GHz to 1.5 GHz.
4. Press the following keys on the analyzer:
Sweep, Sweep (Cont)
BW/Avg, Average (Off)
BW/Avg, Resolution BW (Auto)
Video BW (Auto)
SPAN, 50 kHz
FREQUENCY
5. Press Center Freq. Set the center frequency of the analyzer to the
frequency recorded in the Measured Frequency column (c) in
Table 3-1 for 1.01 GHz to 1.5 GHz.
6. Press the following keys on the analyzer:
BW/Avg, 1 kHz
Video BW Auto Man, 30 Hz
Single
Wait for the sweep to finish.

Table of Contents

Other manuals for Agilent Technologies E4408B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E4408B and is the answer not in the manual?

Agilent Technologies E4408B Specifications

General IconGeneral
BrandAgilent Technologies
ModelE4408B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals