EasyManua.ls Logo

Agilent Technologies E4408B - Page 85

Agilent Technologies E4408B
321 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 3 3-25
Functional Testing
Frequency Readout Accuracy
5. Repeat step 3. If the instrument is functioning correctly, the result
should be between 1.4899990 GHz and 1.4900009 GHz.
NOTE The Frequency Readout Accuracy in now complete. Continue with the
Marker Count Accuracy functional check.

Table of Contents

Other manuals for Agilent Technologies E4408B

Related product manuals