Theory of Operation 4
E4416A/E4417A Power Meters Service Guide 79
A3 Front Panel Assembly
The front panel assembly is made up of a liquid crystal display (LCD), a
keypad and, depending on the power meter option, a power reference cable
assembly and a sensor cable assembly.
There are two inputs to the front panel assembly:
The flex circuit from connector A2J4 of the processor assembly which
controls the keypad.
The ribbon cable from A2J3 of the processor assembly which controls the
LCD display.
The power meter self tests contain display tests. These are
accessed by pressing , ,
, ,
.
System
More
Service
Self Test
Bitmap Displays