Adjustments 3
E4416A/E4417A Power Meters Service Guide 63
Procedure
1 Ensure that the power meter has been powered on for at least 30 minutes
before making any measurements.
2 Connect the equipment as shown in Figure 3- 1. Set up the counter to
measure frequency.
3 Turn the power reference on by pressing:
, on the Agilent E4416A.
, , , on the Agilent E4417A.
4 Observe the reading on the frequency counter.
If it is 50.0 ±0.5 MHz, no adjustment of the power reference oscillator
frequency is necessary. If it is not within these limits, adjust the power
reference oscillator frequency as described in step 5 and step 6.
5 Remove the power meter cover.
6 Adjust A2L9 to obtain a 50.0 ±0.5 MHz indication on the frequency counter.
Refer to Figure 3- 2 for the position of A2L9.
Power Ref On
Cal
More
Power Ref On
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Table of Contents

Other manuals for Agilent Technologies E4417A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E4417A and is the answer not in the manual?

Agilent Technologies E4417A Specifications

General IconGeneral
BrandAgilent Technologies
ModelE4417A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals