2Performance Tests
22 E4416A/E4417A Power Meters Service Guide
Equipment
Power sensor: Agilent E9321A.
2 × Power sensor cable: Agilent E9288A.
30 dB reference attenuator: Agilent 11708A.
Digital voltmeter (DVM): Agilent 3458A.
Breakout box kit: Agilent E9288 opt. K01
(comprises a breakout box (coupler) and a pair of interconnecting red/black
leads).
Test S et up
Figure 2-2 Instrument Accuracy (Normal Path) Test Setup
Breakout box
Digital Voltmeter
+ input
black
red
Power sensor cable
CHANNEL A
Power meter
Power sensor
Power ref
Power sensor
cable
input