2Performance Tests
32 E4416A/E4417A Power Meters Service Guide
Test Procedure 2
Equipment
Range Calibrator: Agilent 11683A Option H01
Power Sensor Cable: Agilent E9288A
DC Calibrator: Datron 4000A
Test S et up
NOTE
This test procedure makes use of a 11683A Range Calibrator fitted with Option H01.
The measurement uncertainty of this test procedure is significantly lower than the
previously described method (refer to Appendix D, Measurement Uncertainty Analysis -
Instrument Accuracy Test for a description of the measurement uncertainty analysis).
As such, this test procedure should be used to ensure the power meter meets its
published specification.
Figure 2-5 Instrument Accuracy (Average Path) Test Setup
Power Meter
Power sensor cable
Range Calibrator
DC Calibrator
CHANNEL A
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Table of Contents

Other manuals for Agilent Technologies E4417A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E4417A and is the answer not in the manual?

Agilent Technologies E4417A Specifications

General IconGeneral
BrandAgilent Technologies
ModelE4417A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals