3 Adjustments
62 E4416A/E4417A Power Meters Service Guide
Power Reference Oscillator Frequency Adjustment
Description
Variable inductor A2L9 is adjusted to set the power reference oscillator output
frequency to 50.0 ±0.5 MHz. This frequency is a supplemental characteristic.
Equipment
Frequency Counter: Agilent 53132A.
NOTE
Adjustment of the power reference oscillator frequency may also affect the output
level of the oscillator. Therefore, after the frequency is adjusted to 50.0 ±0.5 MHz,
check the output level should be checked as described in Power Reference Oscillator
Level Adjustment on page 65. The power reference oscillator frequency is a nominal
specification.
Figure 3-1
Adjustment Setup
Power Meter
Frequency Counter
POWER REF