2Performance Tests
28 E4416A/E4417A Power Meters Service Guide
Instrument Accuracy (Average Path) Test
Specification
Description
The power meter accuracy is verified for various power inputs. There are two
different methods described in this section that can be used to perform this
test. One of these methods has a higher measurement uncertainty value
associated with it.
Test Procedure 1
Equipment
Range calibrator: Agilent 11683A.
Power sensor cable: Agilent E9288A.
Electrical Characteristics Performance Limits (maximum)
Accuracy ±0.5% or 0.02 dB
*
* This performance limit does not include the corresponding sensor power linearity specification.
NOTE
The measurement uncertainty of this test procedure is relatively high at power levels of
30 μW and below, as it uses a standard 11683A Range Calibrator. As such, this test
procedure should be used for indication only.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Table of Contents

Other manuals for Agilent Technologies E4417A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E4417A and is the answer not in the manual?

Agilent Technologies E4417A Specifications

General IconGeneral
BrandAgilent Technologies
ModelE4417A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals