Specifications 1
E4416A/E4417A Power Meters Service Guide 9
Trigger
Channel Functions: A, B, A/B, B/A, AB, BA, and relative.
Storage Registers: 10 instrument states can be saved via Save/Recall
menu.
Predefined setups: For common wireless standards (GSM900, EDGE,
NADC, iDEN, Bluetooth, IS- 95 CDMA, WCDMA
and cdma2000),predefined setups are provided.
Sources: Internal, External TTL, GPIB, and RS232/422.
Time Resolution: 50 ns.
Delay Range: ±1.0 s.
Delay Resolution: 50 ns for delays50 ms otherwise 200 ns.
Hold- off: Range: 1
μ
s to 400 ms
Resolution: 1% of selected value (minimum of
100ns).
Internal Trigger: Range:
20 to +20 dBm
Level Accuracy: ±0.5 dB
Resolution: 0.1 dB
Latency: 500 ns ± 100 ns
Latency is defined as the delay between the
applied RF crossing the trigger level and the
power meter switching into the triggered state.
External Trigger
Range:
High >2.0 V, Low <0.8 V; BNC connector; rising or
falling edge triggered; input impedance >1 kΩ
Trigger Out: Output provides TTL compatible levels (High >2.4
V, Low <0.4 V) and uses a BNC connector.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Table of Contents

Other manuals for Agilent Technologies E4417A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E4417A and is the answer not in the manual?

Agilent Technologies E4417A Specifications

General IconGeneral
BrandAgilent Technologies
ModelE4417A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals