1 Specifications
10 E4416A/E4417A Power Meters Service Guide
Sampling Characteristics
Rear Panel Inputs/Outputs
• Sampling Rate: 20 Msamples/second.
• Sampling Technique: continuous sampling.
• Recorder Output(s): Analog 0 to 1 V, 1 kΩ output impedance, BNC
connector (Two outputs are available on the
E4417A: Channel A and B).
• Remote
Input/Output:
TTL Output: used to signal when measurement has
exceeded a defined limit.
TTL Input: initiates zero and calibration routine
Connector type: RJ- 45 Series shielded modular
jack assembly.
TTL Output: high=4.8 V max, low=0.2 V max.
TTL Input: high=3.5 V min., 5 V max;
low=1 V max, 0.3 V min.
• RS- 232/422
Interface:
Serial interface for communication with an
external controller, Male/Plug 9 pin
D- subminiature connector.
• Trigger In: Accepts a TTL signal for initiating measurements,
BNC connector.
• Trigger Out Outputs a TTL signal for synchronizing with
external equipment, BNC connector.