EasyManuals Logo
Home>Agilent Technologies>Measuring Instruments>E4417A

Agilent Technologies E4417A User Manual

Agilent Technologies E4417A
183 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #31 background imageLoading...
Page #31 background image
1 Specifications
10 E4416A/E4417A Power Meters Service Guide
Sampling Characteristics
Rear Panel Inputs/Outputs
Sampling Rate: 20 Msamples/second.
Sampling Technique: continuous sampling.
Recorder Output(s): Analog 0 to 1 V, 1 kΩ output impedance, BNC
connector (Two outputs are available on the
E4417A: Channel A and B).
Remote
Input/Output:
TTL Output: used to signal when measurement has
exceeded a defined limit.
TTL Input: initiates zero and calibration routine
Connector type: RJ- 45 Series shielded modular
jack assembly.
TTL Output: high=4.8 V max, low=0.2 V max.
TTL Input: high=3.5 V min., 5 V max;
low=1 V max, 0.3 V min.
RS- 232/422
Interface:
Serial interface for communication with an
external controller, Male/Plug 9 pin
D- subminiature connector.
Trigger In: Accepts a TTL signal for initiating measurements,
BNC connector.
Trigger Out Outputs a TTL signal for synchronizing with
external equipment, BNC connector.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Table of Contents

Other manuals for Agilent Technologies E4417A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E4417A and is the answer not in the manual?

Agilent Technologies E4417A Specifications

General IconGeneral
BrandAgilent Technologies
ModelE4417A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals