C Measurement Uncertainty Analysis - Power Reference Level Test
C-142 E4416A/E4417A Power Meters Service Guide
The reference is designed to have a VSWR of less than 1.05.
VSWR < 1.05
Hence,
So u(M
s
) from the equation above equals:
Mismatch Uncertainty U(M
s
) worst case:
Calibration factor
The calibration factors supplied with the power sensor take into account the
performance of the sensor with respect to frequency. This value is of
secondary importance when calculating the overall uncertainty of the 1 mW
reference.
The information that is appropriate is the Uncertainty of the Calibration
Factor. The power sensor must be calibrated such that:
Uncertainty of Calibration Factor < 0.2%
Therefore,
U(CF) = 0.002
P
s
max()P
d
max()
VSWR 1
VSWR 1+
-------------------------
1.05 1
1.05 1+
------------------- 0.0244== ==
uM
s
()
0.0244()0.0244()×
2
--------------------------------------------------
± 0.00042==
UM
s
2P
s
P
d
22 uM
s
()× 0.0012== =
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Table of Contents

Other manuals for Agilent Technologies E4417A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E4417A and is the answer not in the manual?

Agilent Technologies E4417A Specifications

General IconGeneral
BrandAgilent Technologies
ModelE4417A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals