5Replaceable Parts
98 E4416A/E4417A Power Meters Service Guide
NOTE
After replacing a processor board, the display brightness and contrast must be
adjusted. For further information refer to Chapter 3,Display Brightness and Contrast
Adjustment”.
NOTE
Firmware should be downloaded to the instrument after the processor board is
replaced. Refer to Firmware Upgrades on page 94.