C Measurement Uncertainty Analysis - Power Reference Level Test
C-138 E4416A/E4417A Power Meters Service Guide
So the total power incident or measured by the Agilent 432 Power meter is
given by:
The equation above includes the worst- case mismatch magnitude 2P
s
P
d
and is
in the form where various powers can be multiplied by the equation to find
worst- case power levels with respect to mismatch.
P
o
1 P
d
2
()()1 P
s
P
d
±()
2
NOTE
In all cases P
d
is extremely small therefore P
d
²
terms tend to zero leaving.
M
s
Unc erta inty 112P
s
P
d
±()=
NOTE
From the above mismatch uncertainty equation if Pd is zero, for example, the 8478B
sensor is exactly 50 ohms then there is NO mismatch uncertainty. This has not yet been
achieved but efforts will be made in the future to try to achieve as near as possible to a
'zero' level of mismatch uncertainty and hence have negligible mismatch when
calibrating products.