112 Agilent 1200 Series Autosampler Reference Manual
4 Repairing the Autosampler
1 Select “Needle Down” until the needle screw is aligned
with the hole in the safety cover.
2 Remove the sample-loop fitting from the needle fitting.
3 Loosen the fixing screw, and lift out the needle. 4 Insert the new needle. Align the needle in the seat, then
tighten the screw firmly.