Agilent 1200 Series Autosampler Reference Manual 113
Repairing the Autosampler 4
5 Reconnect the sample-loop fitting to the needle fitting. 6 Use “Needle Up” to lift the needle to a position
approximately 2mm above the seat.
7 Ensure the needle is aligned with the seat.
On completion of this procedure:
• Install the front cover.
• Select “End” in the maintenance function “Change
Needle” (see “Change Needle” on page 86).
CellFrame