Agilent 1260 Infinity VWD User Manual 105
Test Functions
8
Intensity Test
Instensity Test with Agilent Lab Advisor
Figure 33 Intensity Test with Agilent Lab Advisor
Intensity Test Failed
Probable cause Suggested actions
1 Empty flow cell
Ensure the flow cell is filled with water.
2 Flow cell windows dirty
Repeat the test with the flow cell removed. If
the test passes, exchange the flow cell
windows.
3 Optics defect
Please contact your Agilent service
representative.
4 Defective lamp or optics.
Exchange the lamp.