66 Agilent 1260 Infinity VWD User Manual
4 Using the Detector
Special Settings of the Detector
Special Settings of the Detector
In this chapter special settings of the detector are described.
Control Settings
Figure 24 Detector control settings
•Lamp: turn on and off of UV-lamp.
•At Power On: automatic lamp-on at
power on.
Error Method: take error method or
current method (in case of an error).
Analog Output Range: can be set to
either 100 mV or 1 V full scale, see
Analog Output Settings” on page 69.
Lamp Type: can be set to either
G1314-60100 (standard VWD lamp) or
2140-0590 (DAD lamps), see also
“Exchanging a Lamp” on page 123.
Automatic Turn On: lamps can be
programmed (detector must be on for
this).
•Help: online help.