Agilent 1260 Infinity VWD User Manual 95
Error Information
7
Detector Error Messages
Ignition Failed
Error ID: 7452
The lamp failed to ignite. The processor monitors the lamp current during
the ignition cycle. If the lamp current does not rise above the lower limit
within 2 – 5 s, the error message is generated.
Probable cause Suggested actions
1 Defective or non-Agilent lamp.
Exchange the lamp.
2 Defective power supply.
Please contact your Agilent service
representative.
3 Defective main board.
Please contact your Agilent service
representative.
Probable cause Suggested actions
1 Lamp disconnected.
Ensure the lamp is connected.
2 Defective or non-Agilent lamp.
Exchange the lamp.
3 Defective power supply.
Please contact your Agilent service
representative.
4 Defective main board.
Please contact your Agilent service
representative.