126 Agilent 1260 Infinity VWD User Manual
9 Maintenance
Exchanging a Flow Cell
Exchanging a Flow Cell
When If an application needs a different type of flow cell or the flow cell needs repair.
Tools required Description
Wrench, 1/4 inch
for capillary connections
Parts required # p/n Description
1 G1314-60186 Standard flow cell 10 mm, 14 µL, 40 bar
(with RFID tag)
OR 1 G1314-60187 Micro flow cell 3 mm, 2 µL, 120 bar
(with RFID tag)
OR 1 G1314-60183 Semi-micro flow cell 6 mm, 5 µL
(with RFID tag)
OR 1 G1314-60182 High pressure flow cell 10 mm, 14 µL, 400 bar
(with RFID tag)
Preparations Turn the lamp OFF.
1 Press the release buttons and remove the front cover to
have access to the lamp area.
2 Disconnect the inlet and outlet capillaries.