Agilent 1260 Infinity VWD User Manual 125
Maintenance
9
Exchanging a Lamp
5
Replace the front cover.
Next Steps:
6 Reset the lamp counter as described in the User Interface
documentation.
7 Turn the lamp ON.
8 Give the lamp more than 10 min to warm-up.
9 Perform “Wavelength Verification-Calibration” on
page 108 to check the correct positioning of the lamp.