182 Agilent 1260 Infinity VWD User Manual
12 Hardware Information
Setting the 8-bit Configuration Switch (without On-board) LAN
Setting the 8-bit Configuration Switch (without On-board) LAN
The 8- bit configuration switch is located at the rear of the module.
This module does not have its own on- board LAN interface. It can be
controlled through the LAN interface of another module, and a CAN
connection to that module.
Figure 54 Configuration switch (settings depend on configured mode)
All modules without on- board LAN:
default should be ALL DIPS DOWN (= best settings)
Bootp mode for LAN and
19200 baud, 8 data bit / 1 stop bit with no parity for RS- 232
DIP 1 DOWN and DIP 2 UP allows special RS- 232 settings
for boot/test modes DIPS 1+2 must be UP plus required mode
Switch settings provide configuration parameters for serial communication
protocol and instrument specific initialization procedures.
NOTE
For normal operation use the default (best) settings.
NOTE
With the introduction of the Agilent 1260 Infinity, all GPIB interfaces have been removed.
The preferred communication is LAN.