24 Agilent 1260 Infinity VWD User Manual
2 Site Requirements and Specifications
Physical Specifications
Physical Specifications
Tabl e 2 Physical Specifications
Type Specification Comments
Weight 11 kg (25 lbs)
Dimensions
(height × width × depth)
140 x 345 x 435 mm
(5.5 x 13.5 x 17 inches)
Line voltage 100 – 240 VAC, ± 10 % Wide-ranging
capability
Line frequency 50 or 60 Hz, ± 5 %
Power consumption 220 VA, 85 W / 290 BTU Maximum
Ambient operating
temperature
0–55 °C (32–131 °F)
Ambient non-operating
temperature
-40 – 70 °C (-40 – 158 °F)
Humidity < 95 % r.h. at 40 °C (104 °F) Non-condensing
Operating altitude Up to 2000 m (6562 ft)
Non-operating altitude Up to 4600 m (15091 ft) For storing the module
Safety standards: IEC, CSA,
UL
Installation category II, Pollution degree 2 For indoor use only.

Table of Contents