132 Agilent 1260 Infinity VWD User Manual
9 Maintenance
Correcting Leaks
Correcting Leaks
1 Remove the front cover.
2 Use tissue to dry the leak sensor area.
3 Observe the capillary connections and the flow cell area for leaks and
correct, if required.
4 Replace the front cover.
Figure 42 Drying the Leak Sensor
When If a leakage has occurred in the flow cell area or at the capillary connections.
Tools required Description
Tissue
Wrench, 1/4 inch
for capillary connections