Agilent 1260 Infinity VWD User Manual 93
Error Information
7
General Error Messages
Open Cover
Error ID: 0205
The top foam has been removed.
The sensor on the main board detects when the top foam is in place. If
the foam is removed during operation, the lamp and grating drive power is
switched off, and the error message is generated.
Probable cause Suggested actions
1 Fan cable disconnected.
Please contact your Agilent service
representative.
2 Defective fan.
Please contact your Agilent service
representative.
3 Defective main board.
Please contact your Agilent service
representative.
Probable cause Suggested actions
1 The top foam was removed during
operation.
Please contact your Agilent service
representative.
2 Foam not activating the sensor.
Please contact your Agilent service
representative.
3 Defective sensor or main board.
Please contact your Agilent service
representative.